Heated Transfer Lines

Uniform heating to safeguard your sample's chemical signature.

Eliminate Cold Spots and Prevent Condensation

Reliable online gas analysis starts with maintaining total sample integrity.

If your gas stream is extracted at elevated temperatures, keeping it fully vaporized on its way to the analyzer is critical.

To prevent target compounds from condensing or adsorbing onto tube walls, using a high-performance heated transfer line is essential.

Chemlys transfer lines are fully configurable to match your application:

Temperature Regulation: Self-regulating or actively controlled options.

Custom Lengths: Tailored to your installation layout.

Internal Tubing Options: 316L stainless steel, SilcoNert® surface treatments, or PTFE.

Valco valve

Self-Regulating Heated Transfer Lines

Operating Range: Optimised for temperatures between 80°C and 180°C

Control & Efficiency: Precise temperature regulation offering the best price-to-performance ratio on the market.

Enhanced Safety: An independent sensor prevents thermal runaway, ensuring total system protection
High-Quality Materials: Internal PTFE liner for maximum chemical inertness.

Simplified Maintenance: The 1/8" maximum diameter stainless steel tube is easily replaceable thanks to a convenient zip-closure sleeve.

Power Supply: 230V standard; can be powered directly by the analyzer for seamless integration.

Valco valve

Temperature Regulated Transfer Lines (Up to 350°C)

Extended Thermal Range: Engineered for demanding applications requiring temperatures up to 350°C.

Precise Control: Includes a dedicated controller to program exact setpoints. Dual independent control loops prevent local hotspots or thermal overshoots.
Versatile Operation: The ideal solution when setpoints must be dynamically adjusted, precisely monitored, or operated above 150 °C.

Enhanced Safety: An independent sensor prevents thermal runaway, ensuring total system protection
Industrial Build: Features a robust stainless steel internal liner for maximum durability at high temperatures.

Simplified Maintenance: The 1/8" maximum diameter stainless steel tube is easily replaceable thanks to a convenient zip-closure sleeve.

Power supply 230V.

You might also be interested in:

Micro GC Fusion

Micro GC Analyzers

A compact and high-speed gas analyzer designed for all your continuous process monitoring needs.

GC Solutions

GC Solutions

Tailor-made gas chromatographs specifically engineered to meet your unique gas analysis requirements.

GC Industrial P5100

Industrial Gas Chromatography

Robust industrial gas chromatographs built for the most demanding process monitoring in harsh industrial environments.